PRODUCTS

NXT Compounding Double HEPA Filter Extraction Hood

MORE INFO

ABOUT

Our NXT Compounding Extraction Hood provides the highest level of operator and environment protection from harmful contaminates, while weighting non-sterile powders and compounds. Quality materials, components and workmanship can be found in this fully loaded NXT series hood. With the addition of the secondary HEPA Filter, it now becomes USP-800 compliant for hazardous drugs intended for non-sterile compounding.

KEY FEATURES

  • Operator Protection
    The airflow is directed away from the operator, into the hood to be filtered. Both the NXT-330 and 430 models have been successfully ASHRAE 110 tested with excellent results.
  • Particulate Filtration
    The Compounding Extraction Hood filters the air through a HEPA Filter rated at 99.999% with the primary filter located at the entry point of the cabinet, allowing all mechanical and electrical components to be protected. The secondary HEPA Filter is located at the top of the hood. This allows for the filter to be scan tested.
  • Independent Operation
    Internal fan(s) direct the air without the need for being ducted to the building system. The double filtered air is vented back into the room, saving otherwise wasted energy.
  • Sturdy Construction
    Constructed from cold-rolled steel with anti-bacterial powered coating, Interior 304 stainless steel with coved corners, two layers of laminated, toughened, anti-UV glass, and an electroplated work surface.

STANDARD FEATURES

  • High Efficiency Particulate Air (HEPA) filter, rated 99.99% efficient at 0.3 microns
  • Power vertical sliding face shield, fully interlocked with both the supply fan(s) and UV lighting
  • UV lighting
  • Energy-efficient, LED task lighting
  • Quick-change, disposable pre-filters, rated MERV 8
  • Side windows made of two layers of laminated, toughened, anti-UV glass
  • Stainless steel rear baffles complete with easy release catches
  • Pre-cut holes for various service outlets
  • Two pass through ports for electrical wires and tubing
  • LCD display for On/Off controls of fan and lights, UV timer, sash height control, duplex and airflow velocity read-out and alarm
  • 15' power cord, 115V
  • All stainless steel baffle system
  • Sample ports for testing primary HEPA Filter and aerosol injection for secondary HEPA

OPTIONAL FEATURES

  • Base frame with either 30" or 35" high work surface
  • ULPA filtration
  • Casters for base frame
  • GFI duplex for electrical outlets
  • Service valves

A leader in Building Clean Environments

We are very pleased to announce that as a result of a recent merger of Microzone Corporation and Design Filtration Inc., our two companies have become one entity under the name of Design Filtration Microzone Inc.

This is a very exciting venture for both companies, for our employees, customers, distributors and our supplier partners. Our mission at Design Filtration Microzone Inc. is to provide an excellent customer service experience with the unsurpassed knowledge base and the high level of expertise that our customers and distributors have come to expect. This joining of forces has resulted in a more comprehensive product offering including several new equipment lines to be announced in the coming months.

location
map-img

86 Harry Douglas Drive.Ottawa, Ontario
K2S 2C7 Canada

Telephone: 613-831-8318

Toll Free: 888-636-8609

Fax: 613-831-8321

Email: info@dfmzgroup.com